AT 058 DPD Neutralising p/n AT058

In Stock

$ 43.50

DESCRIPTION

AT 058 DPD Neutralising - 250 tests

This Palintest chlorine test uses the DPD method developed by Dr A T Palin and now internationally recognised as the standard method of testing for chlorine and other disinfectant residuals. In the Palintest DPD method, the reagents are provided in tablet form for maximum convenience and simplicity of use.

Free chlorine reacts with diethyl-p-phenylene diamine (DPD) in buffered solution to produce a pink coloration. The intensity of the colour is proportional to the free chlorine concentration.

Subsequent addition of excess potassium iodide induces a further reaction with any combined chlorine present. The colour intensity is now proportional to the total chlorine concentration; the increase in intensity represents the combined chlorine concentration. In this way, it is possible to differentiate between free and combined chlorine present in the sample. The colour intensities are measured using a Palintest Photometer.

The DPD Oxystop tablet is added after measurement for free chlorine but before the DPD 3 tablet. It prevents the reaction between shock treatment chemicals and potassium iodide which would give a positive response.